5
•
Installation
UHPLC System
Operating Manual
Page 65
5.5.2
Flow Connections Overview
The picture illustrates the liquid flow path through the system:
Figure 22: Flow connections in the Vanquish system (example)
The Vanquish system has been designed for optimized and simple leak
liquid and waste handling:
•
Leak liquid from the solvent rack, detector, autosampler, and pump
flows through pipes from the modules to the system base drain port.
•
Seal wash and needle wash liquids from the pump and autosampler,
as well as condensate from the autosampler, are also guided to the
system base drain port.
•
Leak liquid from the column compartment flows to the column
compartment drain port.
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...