5
•
Installation
UHPLC System
Operating Manual
Page 55
TIP
Thermo Fisher Scientific recommends using the USB ports only as
described above. If the USB ports are used for any other purpose,
Thermo Fisher Scientific cannot ensure proper functionality.
5.4.2.2
USB Connections
All USB connections require standard USB A-to-B type cable.
Figure 11: USB cable
The flat, rectangular end of the cable is the "A" connector. The smaller,
hexagonal end is the "B" connector.
TIP
•
The USB standard limits the USB cable length to 5 meters. Each USB
device can be separated from the computer or next USB hub by no
more than 5 meters.
•
After connecting the USB cables and before turning on the power to
a device for the first time, verify that the chromatography software
is installed on the computer. When the power is turned on, the
required USB drivers are automatically found and the Windows
operating system can detect the device.
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...