5
•
Installation
UHPLC System
Operating Manual
Page 43
5.3
Site Requirements
The operating environment is important to ensure optimal performance
of the device. This section provides important requirements for the
installation site. Note the following:
•
Make sure that the installation site has enough power outlets for all
devices in the system.
•
Operate the system only under appropriate laboratory conditions
•
For specifications, see
.
5.3.1
Workbench
The Vanquish system is designed to be placed on a workbench. If the
workbench shall carry a complete Vanquish system and possibly other
instruments, the bench must be capable to bear the weight of all devices
and instruments, including solvents.
For dimensions and weight, see
Workbench Requirements
Part
Requirements
Workbench
•
Sturdy
•
A height that ensures convenient access to the
interior of each device in the system.
•
The bench top must be dry, clean, and resistant to
chemicals.
Site for the workbench
The workbench must stand in a secure and level
position that is free of vibrations.
System Clearance Requirements
Side
Requirements
On the sides
•
Allow sufficient free space for electrical
connections and for proper air circulation
•
At least 5 cm of clearance on each side
•
If the system includes a Charger, allow at least 10
cm of clearance on the left side
On the rear
•
Allow sufficient free space for electrical
connections and for proper air circulation
•
At least 15 cm of clearance
Top
At least 30 cm of clearance above the top
TIP
Make sure that the power switch and power cord can be easily reached
at any time.
Содержание thermoscientific Vanquish UHPLC System
Страница 1: ...thermoscientific Vanquish UHPLC System Operating Manual 4820 3601 EN Revision 4 0 December 2019 ...
Страница 4: ...Contacting Us Page 4 UHPLC System Operating Manual ...
Страница 28: ...2 Safety Page 28 UHPLC System Operating Manual ...
Страница 34: ...3 System Overview Page 34 UHPLC System Operating Manual ...
Страница 38: ...4 Unpacking Page 38 UHPLC System Operating Manual ...
Страница 82: ...5 Installation Page 82 UHPLC System Operating Manual ...
Страница 100: ...6 Operation Page 100 UHPLC System Operating Manual ...
Страница 112: ...7 Maintenance and Service Page 112 UHPLC System Operating Manual ...
Страница 128: ...8 Troubleshooting Page 128 UHPLC System Operating Manual ...
Страница 132: ...9 Specifications Page 132 UHPLC System Operating Manual ...
Страница 144: ...11 Appendix Page 144 UHPLC System Operating Manual ...
Страница 148: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...