5
•
Installation
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 69
9. Check the solvent lines across the entire flow path. Make sure that
the lines are not bent, pinched, or squeezed at any point in the flow
path.
10. If applicable, reinstall the seal wash detector assembly.
5.6.6
Installing Solvent Shut Off Valves
These steps apply only when installing a shut off valve
during
initial
setup of the pump. To install a shut off valve
after
initial operation of the
pump, follow the steps in
Installing Solvent Shut Off Valves (
.
When
To shut off the solvent flow through the system at the solvent reservoir,
for example, to prevent solvent from flowing through the system when
you open a flow connection on the low-pressure side
Parts and additional items required
•
Shut off valve, kit containing one shut off valve and fitting
connections
•
Tubing cutter
Preparations
Prepare the pump and follow the steps in
.
Follow these steps
Figure 20: Installing a shut off valve in the solvent line
No.
Description
1
Solvent line
2
Fitting
3
Ferrule
4
Shut off valve
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...