5
•
Installation
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 63
5.6.3.3
Connecting Non-Viper Capillaries
To connect capillaries with conventional fitting connections (using
ferrules) (non-Viper), follow these steps:
1. Insert the capillary into the connection port.
2. Tighten the connection with a wrench (wrench size: 1/4").
NOTICE
Overtightening may damage the connection. With a torque wrench, the
torque should not exceed 3 Nm (for steel capillaries).
3. Check whether the connection leaks. If leakage exists, follow the
steps further down.
Resolving Leakage with Conventional Fitting Connections
1. Remove the capillary.
2. Clean the connection port with a cleaning swab.
3. Reinstall the capillary.
4. If the connection continues to leak, install a new capillary.
NOTICE
To avoid increased dead volume, leakage, or damage to the system,
reuse fittings and ferrules only for the same capillary connection.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...