Leak Detection
Leaks are a potential safety issue.
The leak sensor inside the device monitors the device for liquid leaks
from the flow connections. The liquid is collected in the leak tray and
guided to the drain port. From the drain port, the liquid is discharged to
waste through the drain system of the Vanquish system.
When the leak sensor detects leakage, the status indicators change to
red and beeping starts to alert you. Follow the instructions in this
manual to find and eliminate the source for the leakage.
If leakage is not resolved after 3 minutes, the pump stops the flow.