3
•
Device Overview
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 37
3.6
Mixing System
The pump is shipped with a mixing system, comprising the following
mixers:
•
Capillary mixer
In the capillary mixer, the solvent streams delivered by the pump are
combined and premixed.
•
Static mixer
The static mixer improves the mixing quality.
Mixing with capillary mixer and static mixer results in smoother
baselines.
The pump is shipped with the following mixing system:
Pump
Mixing System
Binary VC-pump
Quaternary pump
Dual pump
Mixing system, volume 400 µL, comprising:
•
Static mixer, volume: 350 µL
•
Capillary mixer, volume: 50 µL
The dual pump has two mixing systems.
Binary VF-pump
Isocratic pump
Mixing system, volume 200 µL, comprising:
•
Static mixer, volume: 150 µL
•
Capillary mixer, volume: 50 µL
With the isocratic pump, the mixing system serves as
inline filter.
TIP
Mixing systems with other volumes are available as an option (see
Gradient Delay Volume, Dwell Volume and Ripple (
).
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...