![Thermo Scientific Surveyor PDA Plus Скачать руководство пользователя страница 132](http://html1.mh-extra.com/html/thermo-scientific/surveyor-pda-plus/surveyor-pda-plus_hardware-manual_1105373132.webp)
A
Firmware
Updating the Firmware
112
Surveyor PDA Plus Detector Hardware Manual
Thermo Scientific
Figure 86.
Surveyor Firmware Upgrade Utility, showing the PDA_Plus_*.bin file selected
3. After you finish upgrading your firmware:
a. Close the Surveyor Firmware Upgrade Utility.
After you close the Surveyor Firmware Upgrade Utility, the connection with the
Surveyor PDA will be terminated automatically.
b. Depress the On/Off switch on the front of the Surveyor PDA Plus Detector to turn
off the power.
c.
Set the rotary switches back to their original settings.
d. Turn the PDA power on to resume normal operation.
Note
You can now safely turn off the power to the detector.
IMPORTANT
It is important to turn the PDA power Off before adjusting the
rotary switches