Routine and Preventive Maintenance
Maintaining the Probe Heater
Thermo Scientific
Surveyor MSQ Plus Hardware Manual
77
Figure 60.
Partially disassembled probe heater
2. Using the parts supplied in the repair kit, rebuild the assembly:
a. Insert a new detent screw insulator into the underside of the probe heater body.
b. Slide the probe screw sleeve over the top of the detent screw insulator.
c. Insert the spring that you removed from the broken assembly into the spring cup.
d. Align the bottom of the spring cup with the probe screw sleeve.
e.
Insert the spring screw into the spring cup and tighten the assembly with a flat-blade
screwdriver.
Spring screw
Spring cup
Probe screw sleeve
Detent screw insulator
Spring