Routine and Preventive Maintenance
Maintaining the APCI Probe
72
Surveyor MSQ Plus Hardware Manual
Thermo Scientific
Figure 54.
Enlarged view of the tip of the APCI probe
Installing the APCI Probe
Y
To install the APCI probe into the probe heater
1. Turn the locking plate of the probe clockwise to the open position, and then carefully
slide the APCI probe into the probe heater. See
.
2. Turn the locking plate counterclockwise to the closed position to lock the probe in place.
.
3. Connect the PEEK fingertight fitting to the APCI probe. See
.
0.41 mm
depth of APCI spacer
0.5 mm
10 mm
CAUTION
Take care when inserting the APCI probe into the probe heater. The capillary
protruding from the end of the probe is fragile and easily damaged. In addition, you can
contaminate the capillary if you let it touch the inner surface of the probe heater as you
insert the probe into the probe heater.