Routine and Preventive Maintenance
Maintaining the ESI Probe
62
Surveyor MSQ Plus Hardware Manual
Thermo Scientific
Figure 41.
ESI ceramic mount and ceramic sleeve
6. Insert a new ESI ceramic sleeve (P/N FM103394) into the ESI ceramic mount. See
7. Place the O-rings in position in the ESI ceramic mount, and then insert the ESI probe
mount into the ESI ceramic mount and ceramic sleeve assembly.
8. Insert the two M3
×
8 cap head stainless steel screws into the ESI probe mount and
tighten with the 2.5 mm Allen key.
9. Reinstall the ESI capillary and reassemble the probe as described in
4.00 mm OD
3.90 mm OD
3.80 mm OD
ESI ceramic mount
ESI ceramic sleeve