4
Start-up
Connecting the USB interface:
4-16
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Connecting the USB interface:
The units come equipped with a USB interface as a standard feature. The connection to the PC is
made using a commercially available USB 1 or USB 2 cable (incubator side: USB male connector of
Type B, PC side: USB male connector of Type A.
The USB interface complies with Standard USB 1.1 and is compatible with Standard USB 2.0 and 3.0
(full speed). Data communication between the PC and incubator across the USB interface requires that
suitable driver software be installed on the PC, unless such software is provided by the currently
installed Windows-version. The installation of the driver software is described in
The bit rate of the interface can be changed within the pre-set Baud rates limits (9.600, 19.200,
38.400, 57.600 Baud). Baud rate adjustment is described in section
“USB interface baud rate setting“
.
Connecting the alarm contact:
Function:
When system errors and failures occur in the temperature or gas control circuits, an alarm message is
issued to the connected alarm/monitoring system. The potential-free contact (1 changeover contact)
has been designed for the following circuits:
Alarm relay:
Note
Skilled work:
Thermo Scientific warrants the operational safety and the operativeness of the device only if
installation and repairs are performed properly.
The connection of the device to an external alarm system must only be carried out by adequately
trained and authorized expert electrical/telecommunication personnel!
Circuit
Mains Voltage
External fusing
Circuits with AC mains
voltage
250 V ~ max.
6 A max.
SELV – circuits (cf.
VDE 0100,
part 410)
25 V ~
2 A max.
60 V =
1 A max.
SELV E – circuits (cf.
VDE 0100,
part 410)
50 V ~
1 A max.
120 V =
0.5 A max.
Operating states
Contact 3 - 2
Contact 3 - 1
Operating state power
failure “OFF”
X
O
Operating state power
failure “ON”
..% o
X
Error power failure
“OFF”
..% o
X
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...