Description of the device
Door switch
3-10
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
N
2
supply:
If the oxygen content during operation is to be lowered to less than 21% (air oxygen content), the
workspace is supplied with nitrogen. The O
2
concentration in the workspace atmosphere can be
controlled depending on sensor technology.
Door switch
Figure 3-6.
Door switch
A door switch is installed at the upper edge of the workspace opening. When the door switch is
activated by opening the glass door, the gas supply and workspace heating are suspended. The display
shows a corresponding message.
An alarm sounds, if the door is open longer than 30 seconds. If the door remains open for more than
10 minutes, an audible alarm sounds and the alarm relay responds.
The outer door can only be closed after the glass door has been latched properly.
.
Note
Versions with gas-tight screen:
For devices with the optional gas-tight screen, the door switch function described above is activated
when the outer door is opened.
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...