9
Maintenance
Temperature Calibration Procedure
9-4
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Figure 9-2.
Temperature display field and temperature selection menu
2. To exit the Temperature menu:
• Press the
End
key [3].
3. To enter the Calibration submenu:
• Press the
Calibration
key.
Figure 9-3.
Temperature Calibration Procedure
4. Enter the measured value (target value):
The targeting value can be increased or reduced in increments; if you keep the + key or the – key
depressed, the function switches to a rapid increase/reduction; after approx. 3 seconds, another
increase/reduction occurs.
To increase the targeting value:
• Press the + key.
To reduce the set value:
• Press the - key.
5. Accept and save the target value:
• Press the Enter key.
• Press the Save key.
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...