Handling and control
User configuration
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
6-33
Figure 6-37.
Alarm relay setting
1. To toggle between two states:
• Press the + key.
or
• Press the - key.
2. To accept and save the change:
• Press the
Enter
key.
• The system returns to the
Options
menu.
Low humidity setting
If condensation occurs on the culture containers due to high relative humidity, the humidity in the
workspace can be set to a lower level. At the factory, the device control is preset to high humidity
(approx. 93% relative humidity.
The relative humidity in the workspace is lowered from approx. 93% to approx. 90%.
The modification requires an extended adaption phase. To ensure that it effectively prevents dew
formation on culture containers, it must be used as a permanent setting.
Reducing humidity in the workspace
1. Press the
Menu
key.
2. Select the menu command
Options
.
• The selection dialog shown in
3. Select the menu command Low humidity.
is displayed.
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...