8  |   Operation

SpeedVac™ SPD210

c. After the rotor stops spinning, the cover unlocks and 

the lid locked indicator is turned off. The display 
reverts to last set parameters. 

d. Open the cover and remove samples.

GENERAL: During the run, display shows actual parameters. 
To check set parameters, press 

View

 button and 

Select

 

button. The display will revert temporarily to the set points for 
approximately 5 seconds.

Drying Rate

Thermal energy can be applied to the concentrator chamber 
to counteract the cooling effect of evaporation, maintain the 
samples in the liquid state, and accelerate the concentration 
run. Select the highest drying rate if you desire accelerated 
drying. To ensure against over drying of samples, and 
possible denaturing, it is recommended that the “Heater” 
timer be set for only a portion of the total run time (e.g., 50%-
75% of total run time).

Exporting Live Run Data

The unit is equipped with a USB port to enable 
communication with an external computer. The USB port is 
located on the rear panel of the unit and is connected to a 
computer via a standard USB type A to B cable.

Figure 2. USB Port location

When a run is ongoing, the unit exports the following 
information via the USB port every minute.

The live temperature applied to the samples chamber

The remaining heat time for the run (“CCC” indicates 
continuous heating)

The remaining or elapsed run time, depending on the run 
type (manual or automatic)

The data are sent at a fixed baud rate of 115200 and are 
comma-separated as shown in the format: 

<temperature>, 

<heat time>, <run time>.

The temperature is represented as a 2-digit integer in °C

The heat time is expressed as a 3-digit integer in minutes 
(or “CCC” for continuous heating)

The runtime is expressed as a 3-digit integer in minutes.

Note: 

The following section details the step-by-step 

procedures for configuring Microsoft

®

 HyperTerminal running 

on a host computer using Windows

®

 XP. These instructions 

may need to be modified to be used with a different terminal 
emulator program and/or operating system. Contact 
Technical Services if further assistance is required.

HyperTerminal 
Configuration

1. Power up the host computer and close any running 

applications.

2. Open the HyperTerminal application by clicking on 

Start

 \ 

“Programs” \ “Accessories” \”Communications” \ 
“HyperTerminal.”

3. In the “Connection Description” box, enter the name 

“SPD210” and choose an icon and click 

OK

.

4. Connect to the virtual COM port that is linked with the 

SpeedVac™ concentrator attached to the computer 
(USB virtual com port).

5. In the “COM Port Properties” box \ “Port Settings” folder 

select the following options:

Bits per second: _ 115200

Data bits: _ 8

Parity: _ None

Stop bits: _ 1

Flow control _ None

After verifying the above settings, click 

OK

.

6. In the main dialog box click on 

File

 \ 

Save

.

CAUTION: 

 USB lead length less than 3 meters 

shall be used to connect to the computer

In case of a communication failure between the 
monitoring program of the computer and the 
product, user can re-establish the connection by 
disconnecting and reconnecting the USB cable.

Содержание SpeedVac SPD210

Страница 1: ...Installation and Operation 80302131 Revision A September 2018 SpeedVac SPD210 Vacuum Concentrator ...

Страница 2: ...rmo Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited except with the written authorization of Thermo Fisher Scientific Inc The contents of this document are subject to change without notice All technical info...

Страница 3: ...ion 4 Contents 4 Installation of SPD210 4 Operation 5 Description Of Control Panel 5 Manual Run 7 Auto Run 7 Drying Rate 8 Exporting Live Run Data 8 HyperTerminal Configuration 8 Lid Stay Mechanism 9 Other Topics 9 Rotors and Rotor Blocks 10 Accessories 11 Appendix 1 12 Additional SPD210 System Setup 12 Appendix 2 13 Warranty 14 ...

Страница 4: ...is completed with the addition of a vapor condensation trap available separately for virtually 100 solvent recovery This results in a thorough evaporation process in a safe environment for both the sample and the user The unit features an advanced front panel with dual timer for automatic or manual control of run conditions it also features three programs that are modifiable which enable user to r...

Страница 5: ...dicates extreme low temperatures and high risk of frostbite Do not touch bare metal or samples with unprotected body parts This symbol indicates possible pinch points which may cause personal injury This symbol indicates a need to use gloves during the indicated procedures If performing decontamination procedures use chemically resistant gloves Before installing using or maintaining this product p...

Страница 6: ...per Installation Categories II Operative Power 115 VAC 60 Hz 8 amps 230 VAC 50 Hz 6 amps Vacuum Chamber TEFLON coated chemical resistant aluminum casting Cover Glass cover includes a safety interlock Induction Motor Maintenance free Bearings Standard Chamber temperature 35 C to 65 C 5 C increments Weight 117 lbs 53 kg Dimensions W x D x H 19 in x 26 in x 18 in 48 cm x 66 cm x 46 cm Fuse 8 A 250 VA...

Страница 7: ...units configured for 230 VAC 50 Hz should be plugged into a circuit rated for at least 6 amps Note Do not use any detachable power cord that is not adequately rated for the unit Installation of SPD210 Figure 1 Rear View of System Table 1 Contents Qty Description 1 Thermo Scientific Savant SpeedVac SPD210 Concentrator 1 Cover Lock Emergency Release Tool 1 Line Cord 1 To prevent damage to pump due t...

Страница 8: ...the pin with the groove Secure the assembly by screwing the retaining knob into the drive shaft above the rotor Tighten it firmly but not excessively Description Of Control Panel 1 Lid locked Indicates if the lid of the concentrator is locked or not The indicator will be illuminated when locked 2 Program buttons 3 programs that are modifiable Loading a program is achieved by pressing the correspon...

Страница 9: ...parameter 9 View During a run pressing view shows the set run parameters When no run is ongoing pressing view displays the actual live parameters The time display shows either 0 00 or the end status of the last executed run e g End Err 10 Manual run Starts a Manual run based on the parameters loaded in the screen use manual run function if you do not want to set a fixed run time 11 Auto run Starts...

Страница 10: ...es will click allowing air to bleed into the chamber 11 After the rotor stops spinning the cover lock disengages and the lid locked indicator is turned off The display reverts to last set parameters 12 Open the cover and remove the samples GENERAL During the run display shows actual parameters To check set parameters press View button and Select button The display will revert temporarily to the se...

Страница 11: ...he data are sent at a fixed baud rate of 115200 and are comma separated as shown in the format temperature heat time run time The temperature is represented as a 2 digit integer in C The heat time is expressed as a 3 digit integer in minutes or CCC for continuous heating The runtime is expressed as a 3 digit integer in minutes Note The following section details the step by step procedures for conf...

Страница 12: ...ignal the user with the Lid message on the display and an error tone Otherwise during a run the user cannot open the cover due to vacuum in the concentration chamber The cover is locked down at all times during a run and whenever power to the unit is interrupted The cover lock is an additional safety feature that reduces the risk of injury or damage while rotor is spinning Never bypass the cover l...

Страница 13: ...itable measuring and auxiliary equipment as well as detailed service instruction Wear gloves eye protection masks and lab coats while working on the system WARNING Burns due to hot device parts When the concentrator is set to a temperature of 60 C and above the rotor and the sample tubes will be hot do not touch the wall of the rotor chamber use suitable auxiliary to remove the sample tubes Type D...

Страница 14: ...x45 mm 1 dram 24 3 RBA24 15 45 Round or Flat Bottom Flask 2 50 ml each flask RBA2 FLSK 100 1 175 ml each flask RBA1 FLSK 250 Part No Description GF4000 Glass condensation flask 4 liter 121 6162 00 White Foam Insulating ring for coverglass trap SCC1 CryoCool heat transfer fluid 1 liter RAD200 Rotor Adapter for the SUMAX400 Upper Magnet assembly ANT100 Post trap assembly for Oligo Preps Used on UVS8...

Страница 15: ...stem Figure 6 Component High Vacuum System SpeedVac Concentrator SPD210 Vacuum Tubing To Pump UVS450 ANT100 Exhaust recovery Post Trap Bleeder Valve Universal Vacuum Source SpeedVac Concentrator SPD210 Flexible Vacuum Tubing RVT450 Cold Trap SCT120 DVG50 VLP120 Vacuum Pump Chemical Trap Bleeder Valve ...

Страница 16: ...es with methanol to remove materials that inhibit or prevent evaporation of aqueous samples 2 TO KEEP VACUUM PUMP OIL CLEAN TRY THE FOLLOWING Empty solvent s collected in the Refrigerated Vapor Trap as frequently as possible Change the oil in the vacuum pump when it becomes brown in color or smells of solvents being evaporated Attach a Thermo Scientific VPOF110 Vacuum Pump Oil Filter Recirculator ...

Страница 17: ...01450 Power switch Molveno SX8231881100F000 1 14 80301442 MAIN CONTROL BOARD 1 15 80302069 EJOT WN1451 K40 LENGTH 8mm TORX KN6038 4 16 80302062 TORX PAN HEAD 6 32 X 3 8in 6 17 80302061 TORX PAN HEAD 6 32 X 1 4in 20 18 M50 0130 00 NUT HEX KEPS 832 SS 4 19 80302072 Torx 8 32 x 1 2 Flathead screw 4 20 80302172 TORX FLAT HEAD SCREW 6 32x5 8in 2 21 80301452 USB CABLE B MALE TO B FEMALE PANEL MOUNT 1 22...

Страница 18: ...4 M60 0015 01 TUBULAR HEATER 120V 1 5 80301582 MOTOR DRIVE MAGNET ASSY 115V 1 6 80302067 TORX PAN HEAD SCREW 8 32x5 16 6 7 80302061 TORX PAN HEAD 6 32 X 1 4in 4 A B TITLE 1 2 SolidWorks 2007 10 25 C D E F E D C B A F 3 4 5 6 7 8 9 9 8 7 6 5 4 3 2 1 A B TITLE 1 2 SolidWorks 2007 10 25 C D E F E D C B A F 3 4 5 6 7 8 9 9 8 7 6 5 4 3 2 1 ASSEMBLY CHAMBER HEATER SPD210 115 ...

Страница 19: ... SX8231881100F000 1 14 80301442 MAIN CONTROL BOARD 1 15 80302069 EJOT WN1451 K40 LENGTH 8mm TORX KN6038 4 16 80302066 TORX FLAT HEAD SCREW 6 32x5 16in 2 17 80302062 TORX PAN HEAD 6 32 X 3 8in 6 18 80302061 TORX PAN HEAD 6 32 X 1 4in 20 19 M50 0130 00 NUT HEX KEPS 832 SS 4 20 80302072 Torx 8 32 x 1 2 Flathead screw 4 21 80302172 TORX FLAT HEAD SCREW 6 32x5 8in 2 22 80301452 USB CABLE B MALE TO B FE...

Страница 20: ... 1 4 M60 0015 02 TUBULAR HEATER 240V 1 5 80301505 MOTOR DRIVE MAGNET ASSY 1 6 80302067 TORX PAN HEAD SCREW 8 32x5 16 6 7 80302061 TORX PAN HEAD 6 32 X 1 4in 4 A B TITLE 1 2 SolidWorks 2007 10 25 C D E F E D C B A F 3 4 5 6 7 8 9 9 8 7 6 5 4 3 2 1 A B TITLE 1 2 SolidWorks 2007 10 25 C D E F E D C B A F 3 4 5 6 7 8 9 9 8 7 6 5 4 3 2 1 ASSEMBLY CHAMBER HEATER SPD210 230 ...

Страница 21: ... 80301449 DAMPER 3M SJ57B1 12 7x1 3 mm 2 5 80301482 Glass cover 1 6 80302033 NUT BRACKET ASSY 2 7 80302022 HINGE SOUTHCO N6 4E 422 20 2 8 80302072 Torx 8 32 x 1 2 Flathead screw 4 9 M90 0050 01 ADHESIVE FOAM STRIP 1 16 X 1 8 x 3 inch 10 80302066 TORX FLAT HEAD SCREW 6 32x5 16in 8 11 80302082 TORX PAN HEAD 6 32 X 5 16in 2 A B TITLE 1 2 SolidWorks 2007 10 25 C D E F E D C B A F 3 4 5 6 7 8 9 9 8 7 6...

Страница 22: ...MA number issued by Thermo Costs of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this manual When returning apparatus that may contain hazardous material you must pack and label them following U S Department of Transportation DOT regulations applying to transportation of hazardous materials Your shipping documents must also meet DOT regulat...

Страница 23: ... 30 France 33 2 2803 2180 33 2 2803 2180 Germany 0800 1 536 376 49 6184 90 6000 0800 1 536 376 Italy 32 02 95059 552 39 02 95059 552 432 254 375 Netherlands 31 76 579 55 55 31 76 571 4440 Nordic Baltic CIS 358 9 329 10200 358 9 329 100 Russia 7 812 703 4215 7 812 703 4215 Spain Portugal 34 93 223 09 18 34 93 223 09 18 Switzerland 41 44 454 12 22 41 44 454 12 12 UK Ireland 44 870 609 9203 44 870 60...

Отзывы: