Preface
iv
75006480 Diagnostik
TM
Rotor
Thermo Scientific
Precautions
In order to ensure safe operation of the diagnostic rack rotor, the following general safety regulations
must be followed:
•
Do not remove the magnet at the rotor bottom.
•
Do not use rotors which show any signs of corrosion and/or cracks.
•
Use only with rotors which have been loaded properly.
•
Never overload the rotor.
•
Operate the rotor always with the lid closed.
•
Use only accessories which have been approved by Thermo Fisher Scientific. Exceptions to this
rule are commercially available glass or plastic centrifuge tubes, provided they have been approved
for the speed or the RCF value of the rotor.
•
Please observe the safety instructions.
Please pay particular attention to the following aspects:
•
Rotor installation: Check that the rotor is locked properly into place before operating the
centrifuge.
•
Always balance the samples.
Maximum sample density at maximum speed: 1.2
g
cm
3
-----------
Содержание Sorvall RC-4
Страница 4: ...C Contents ii 75006480 DiagnostikTM Rotor Thermo Scientific ...
Страница 7: ...Thermo Scientific 75006480 DiagnostikTM Rotor 1 1 1 Rotor Data Contents Technical Data on page 1 2 ...
Страница 11: ...Thermo Scientific 75006480 DiagnostikTM Rotor 2 1 2 Accessories Contents Accessory on page 2 2 ...
Страница 17: ...4 Rotor Loading Improper Loading Thermo Scientific 75006480 DiagnostikTM Rotor 4 3 Improper Loading ...
Страница 18: ...4 Rotor Loading Improper Loading 4 4 75006480 DiagnostikTM Rotor Thermo Scientific ...
Страница 24: ...5 Maintenance and Care Service of Thermo Fisher Scientific 5 6 75006480 DiagnostikTM Rotor Thermo Scientific ...
Страница 33: ......