Operation
3. 5. Entering Centrifugation Parameters
NOTICE
Due to limited display digits there is a need to round the values. The direct comparison
between the two values speed and RCF is therefore restricted.
3. 5. 1. Selecting Speed or RCF-Value
Press the
TOGGLE
button to switch between the two modes.
When the lower indicator is lit the display shows the speed.
When the upper indicator is lit the display shows the RCF value.
3. 5. 2. Pre-selecting Speed
1. Enter the desired value by pressing up and down
ARROW
buttons
repeatedly until
the desired value shows. You can adjust the speed in 100 rpm increments.
2. Press
START
to confirm the pre-selected value.
If you do not press any button, the display flashes for a few seconds. The new
preselected value is now stored and the display shows the actual value.
NOTICE
The centrifuge speed can be set to a minimum of 300 rpm. The maximum speed depends
on the centrifuge variant.
NOTICE
Avoid if possible speed ranges close to natural resonances of the system. Runs at
resonance speeds may cause vibration and have an adverse effect on the quality of separation.