USB Import/Export
19 of 73
3
USB Import/Export
Run logs, rotor logs, programs and users can be exported onto a USB stick.
Programs and users can also be imported. Only data that has been exported using the same language setting and
the same profile set can be imported.
The file format for export is CSV. Data can be edited in a spreadsheet program such as Microsoft Excel™. The run
log is also exported in PDF format.
The USB stick used for import and export must use the FAT32 file system.
If the centrifuge is connected to a network, run logs can also be downloaded from it as PDF files via http via the
centrifuge's IP address and port '801'. Thus, if for example the IP address assigned to the centrifuge is
'192.168.0.1', the address for download is 'http://192.168.0.1:801'.
If an unknown rotor has been inserted into the centrifuge, the user is prompted to import the relevant rotor data. The
rotor is supplied with a suitable USB stick for this purpose. Use this stick to load the rotor data into the centrifuge.