![Thermo Scientific SOLA II Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/thermo-scientific/sola-ii/sola-ii_user-manual_1105272042.webp)
Installation
DCS & External Connections
3-4 SOLA II User Guide
Thermo Fisher Scientific
The following inputs are available for controlling the analyzer from a DCS or
other external source. Refer to the wiring diagrams shipped with your
instrument.
λ
Remote Stream Control: A manual switch or DCS output signal can be
used to select the following stream operation when the multiple stream
option is installed. For this control to function, enable the Remote Control
function the
Table 3–1.
Stream
Select 0
Stream
Select 1
Action
Close*
Open
Analyzes only stream 1
Open
Close*
Analyzes only stream 2
Close*
Close*
Analyzer alternates between stream 1 and stream 2 at the
time intervals specified in the
parameters (Chapter 6)
Open
Open
Analyzer operation for streams is performed as specified in
the
*Close = Shorted to associated ground terminal
λ
Remote Measurement Range Control: To switch to high measurement
range, short the Remote Range connection to the Remote Range common
terminal. To switch to low measurement range, open the connection
across the Remote Suspend connectors. The
must be enabled (“Analyzer Setup”, Chapter 6).
λ
Remote:
Autocal/Validation: The Autocal/Validation function can be initiated by
closing the contact.
Remote Suspend: To suspend analysis, short the Remote Suspend
connection to the Remote Suspend common terminal. To restart analysis,
open the connection across the Remote Suspend connectors. The
must be enabled (“Analyzer Setup”, Chapter 6).
Analog inputs are available for connecting densitometers used to correct each
stream measurement based on the density of the sample.
Optional Analog Inputs
Содержание SOLA II
Страница 1: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB Part of Thermo Fisher Scientific...
Страница 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 16: ......
Страница 38: ......
Страница 50: ......
Страница 84: ......
Страница 92: ......
Страница 150: ......
Страница 159: ...TCP IP Bridge The Browser Interface Thermo Fisher Scientific SOLA II User Guide D 9 Figure D 17 Sample results...
Страница 160: ......
Страница 168: ......
Страница 186: ...Software Menu Tree H 2 SOLA II User Guide Thermo Fisher Scientific Figure H 2 Diagnostics subgroups...
Страница 188: ......
Страница 194: ......
Страница 198: ......
Страница 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...