Chapter 8: Maintenance & Troubleshooting
Safety Precautions
Caution
Some internal components can be damaged by small amounts of static
electricity. Take appropriate precautions (use a properly grounded antistatic
wrist strap) when handling electronic boards and components.
To avoid damaging internal components, follow these precautions when
performing any service procedure:
Wear an antistatic wrist strap that is properly connected to earth ground. If an
antistatic wrist strap is not available, be sure to touch a grounded metal object
before touching any internal components.
Handle all printed circuit boards by the edges.
Carefully observe the instructions in each procedure.
▲
Maintenance
Schedule
Table 8-1. Maintenance schedule
Frequency
Tasks
Monthly
Calibrate the analyzer (see
).
Every three months
Replace the injection valve rotors according to
Every six months
Visually inspect and clean the instrument.
Check the instrument flow rates as explained later in this chapter.
Replace the valve body according to
Annually
Test the instrument for internal leaks as explained later in this chapter.
Every 18 months
Replace the pyrolyzer heater as explained later in this chapter.
Visual Inspection &
Cleaning
The analyzer should be inspected occasionally for obvious visible defects, such
as loose connectors, loose fittings, cracked or clogged Teflon lines, and excessive
dust or dirt accumulation. Dust and dirt can accumulate in the instrument and
can cause overheating or component failure. Dirt on the components prevents
efficient heat dissipation and may provide conducting paths for electricity.
Warning
Remove all instrument power before cleaning electronics.
▲
The best way to clean the inside of the instrument is to first carefully vacuum
all accessible areas and then blow away the remaining dust with low-pressure
compressed air. Use a soft paintbrush or cloth to remove stubborn dirt.
Thermo Scientific
SOLA II Flare User
Manual
8-1
Содержание SOLA II Flare
Страница 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Страница 2: ......
Страница 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 42: ...Chapter 3 Start Up Shutdown Thermo Scientific SOLA II Flare User Manual 3 6 Figure 3 4 Diagnostics Menu...
Страница 44: ......
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 80: ......
Страница 104: ......
Страница 122: ......
Страница 136: ......
Страница 148: ......
Страница 158: ......
Страница 160: ......
Страница 162: ......