Thermo Fisher Scientific
SGD-O User Guide
E-1
Appendix E
Commonly Used Parameter &
Command DACs
The following tables provide regularly used parameter and command
DACs.
Table E–1.
Common parameter DACs
Parameter Description
DAC
Notes
Snapshot of detector signal
112003
0 to 65540 corresponds to 0 to10 V input
Standardization value
128003
0 to 65540
Standardization value corrected for
source decay
121003
Snapshot of channel 1
116013
Snapshot of channel 2
116023
Time constant
007004
128 seconds (8 seconds recommended)
Slope correction factor (labeled
“slope”)
049003
1.000
Calibration density
024003
Top of span (reading at 20 mA)
(Solids added / Volume of carrier)
114013
2500 g/L (metric)
30 lb/gal (English)
Bottom of span (reading at 4 mA)
115013
0.00
Pipe internal diameter
048003
inches
Cal/Ref ratio
086003
1.0
Solids density
083003
2.65 g/cm
3
Carrier density
085003
0.9982 g/cm
3
Carrier attenuation coefficient
019003
0.086 cm
2
/g
Solids attenuation coefficient
020003
0.077 cm
2
/g
Weeks since last reference
054003
Enable less than 0 and greater than
100%
043030
Enable phantom measurements (HV
control)
045020
Содержание SGD-O
Страница 2: ......
Страница 3: ...SGD O Non Contacting Density Gauge For Fracturing Cementing User Guide P N 717910 Revision D...
Страница 4: ......
Страница 6: ...This page intentionally left blank...
Страница 8: ...This page intentionally left blank...
Страница 36: ...This page intentionally left blank...
Страница 66: ...This page intentionally left blank...
Страница 70: ...This page intentionally left blank...
Страница 72: ...This page intentionally left blank...