The Current Output
Clear All Holds
6-2
SGD-O User Guide
Thermo Fisher Scientific
Use the Clear All Holds command (DAC
9
) to return the current output
to its normal operation.
The output of the detector (input to the transmitter) is a voltage signal
0–10 Vdc maximum. Read the signal (DAC
112003
). The value is
displayed in scientific notation and represents a range of 0 to 10 volts with
the numbers 0 to 6.554E4 (65540).
Clear All Holds
Check the Input
Signal