Getting Started
Interface
2-2
SGD-O User Guide
Thermo Fisher Scientific
This section will show you how to change a parameter using the direct
access method. In this example, change the value of the time constant. The
DAC for the time constant parameter is
007004
.
1.
Press:
EXIT SETUP, 007004, down arrow
.
2.
To leave the value as is, press
EXIT SETUP
, and the gauge will return
to its normal display mode.
To change the value, enter the new value, and press the
down arrow,
EXIT SETUP.
The gauge will begin using the new value immediately
and will return to normal display mode.
This section will show you how to issue commands using the direct access
method. In this example, force the current output to 20 mA. The
command code is
5
.
1.
Press:
EXIT SETUP, 5, down arrow
.
2.
The display shows two options. Press the
left arrow
to reject the
command or the
right arrow
to execute it.
Changing or Viewing
Parameters
Issuing Commands
Содержание SGD-O
Страница 2: ......
Страница 3: ...SGD O Non Contacting Density Gauge For Fracturing Cementing User Guide P N 717910 Revision D...
Страница 4: ......
Страница 6: ...This page intentionally left blank...
Страница 8: ...This page intentionally left blank...
Страница 36: ...This page intentionally left blank...
Страница 66: ...This page intentionally left blank...
Страница 70: ...This page intentionally left blank...
Страница 72: ...This page intentionally left blank...