2-18
User Guide for Sarasota ID900, FD900, & PD900 Gas Density Meters
Thermo Fisher Scientific
Installation
Electrical Data
F option
At the amplifier terminals 1 & 2: Ui = 30 V, li = 100 mA, Pi = 0.7 W,
Ci=0.01 µF, Li=0
At the PRT terminals W, X, Y, & Z: Ui = 10 V, li = 300 mA per
terminal, Pi = 0.5 W, Ci=0.01µF, Li=0
H option
At the amplifier terminal pairs 1 & 2, 3 & 4, 5 & 6: Ui = 28.5 V,
li = 100 mA, Pi = 0.7 W, Ci=0.01µF, Li=0
(Terminals 5 & 7 are linked internally)
Terminals 7 to 8: Uo = Ui, lo = li, Po = Pi,
Ci=0.01µF, Li=0
Specific Conditions of Use:
1.
The enclosure may be constructed from aluminum alloy and given a protective paint
finish, however, care should be taken to protect it from impact and friction if located in
Zone 0. The painted enclosure constitutes a potential electrostatic charge hazard and
must not be rubbed with a dry cloth or cleaned with solvents.
2.
When the equipment is installed, care must be taken to ensure the maximum
ambient temperature of the electronics housing is not affected by the specified process
temperature of the connected sensor assembly.
Содержание Sarasota FD900
Страница 2: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank...
Страница 8: ...This page intentionally left blank...
Страница 12: ...This page intentionally left blank...
Страница 36: ......
Страница 42: ...This page intentionally left blank...
Страница 46: ...This page intentionally left blank...
Страница 54: ...This page intentionally left blank...
Страница 58: ...This page intentionally left blank...
Страница 68: ...This page intentionally left blank...
Страница 88: ...This page intentionally left blank...
Страница 89: ......