Maintenance and Trouble Shooting
Thermo Fisher Scientific
SAM12 User Manual 11-9
These items are unlikely to fail or need regular servicing.
However, in the event of failure, their removal and replacement
is obvious. See
Recommended Spares List (page 12-1)
for
component details.
WARNING: Unplug the mains lead and disconnect the
battery to avoid the SAM12 switching on inadvertently as the
start switch connector is unplugged.
Detector Removal and Replacement
NOTE: The detectors are light sensitive and, because of this,
are contained in a light-tight enclosure. Care must be exercised
at all times not to damage the enclosure as subsequent damage
to the photomultiplier tube may result due to light leaks.
Disconnect the mains supply. Ensure the key-switch on the
rear panel is in the OFF position.
Open the door and remove the screws retaining the stainless
steel cubicle Liner to the Main Cabinet frame.
Removing the Stainless Steel Liner from the Cabinet to exposes
the detectors.
Undo the wing-nuts that hold the detectors in place and slide
the detector to the rear of the SAM12.
When space allows disconnect the detector coaxial cable.
The detectors are held in place by 4 lugs, when the detector is
free of the lugs it may be lifted out.
TAKE CARE THAT THE DETECTOR DOES NOT FALL!!
Replacement of the Detectors follows the removal procedure in
reverse. Again care must be exercised not to break the light-
tight seals.
Once the detectors are in place, refit the stainless steel liner.
Replacement detectors or detectors with replacement PMT’s
will need to be set for optimum performance with the Isotopes
of interest. Before determining HV’s, the Single Channel
Analyser Thresholds must be reset to their default settings or
erroneous results will be obtained (see
Thresholds (page 5-28)
).
Once this is done, select the detector operating point using the
HV scan procedure detailed in
Selection of Detector Operating
Parameters (page 9-3)
.
Start Switch, Keyswitch
and Loudspeaker -
Removal & Replacement
Removal
Detector Replacement
Setting up Replacement
Detectors for Use
Содержание SAM12
Страница 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Страница 2: ......
Страница 4: ...This manual was produced using ComponentOne Doc To Help...
Страница 10: ......
Страница 18: ......
Страница 20: ......
Страница 34: ......
Страница 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Страница 130: ......
Страница 134: ......
Страница 144: ......
Страница 168: ......