QuickStart Doc No. 065516-01
Page 1 of 2
January 2013
CHARGE DETECTOR CELL (QDC 300)
Quick Start
(0.4 mm) P/N: 078841
To ensure proper operation of the QDC 300 follow the instructions below. Refer to the ICS 4000 manual and the
QDC 300 manual for detailed operation guidelines.
Do not attempt to disassemble the QDC 300; it may result in irreversible
damage.
WARNING
!
The QDC 300 should be installed after the Conductivity (CD) background on
the system is
≤
2 µS/cm.
1.
Hydration and start up of QDC
The QDC 300 is shipped dry and should be hydrated prior to use. This process ensures that the QDC 300
membranes are fully hydrated and ready for operation. Figure 1A shows the top view of the QDC 300 as
shipped with the “ELUENT IN” line plumbed to the “REGEN OUT” port of the QDC 300. Figure 1B shows the
bottom view of the QDC 300 with the ball stud and connector required for installing the QDC 300.
Figure 1A: Top View of the QDC 300 as Shipped
Figure 1B: Bottom view of the QDC 300
NOTE
!