Thermo Scientific PowerEase PS0300 Скачать руководство пользователя страница 7

PowerEase

®

 300W Power Supply User Guide

  

 

 7

 

 

1B

Methods 

12B

Getting started 

31B

Install the 

PowerEase

®

 300W 

Power Supply 

1.

 

Check the label located at the bottom of the power supply to ensure that the unit is compatible 
with locally provided voltage. 

2.

 

Place the PowerEase

®

 300W Power Supply on a level laboratory bench. Keep the area around 

the power supply clear to ensure proper ventilation of the unit. 

3.

 

For your safety: Position the unit properly such that the On-Off switch and the AC inlet 
located on the rear of the unit are easily accessible. 

4.

 

Ensure the AC power switch is in the Off position. 

5.

 

Attach the power cord to the AC inlet. Use only properly grounded AC outlets and power 
cords. 

6.

 

Connect the leads from the electrophoresis unit; insert the positive lead (+) into the positive 
(red) output jack, and the negative lead (–) into the negative (black) output jack. 

13B

Important guidelines 

32B

Introduction 

The important guidelines for operating the PowerEase

®

 300W Power Supply are provided in this 

section. We recommend that you carefully review these guidelines before operating the instrument. 

Important: 

For best results, do 

NOT 

use the PowerEase

®

 300W Power Supply at its maximum 

electrical load limits. Variations in buffer conditions can result in exceeding the power supply’s 
maximum voltage, current, or power output capacity and produce undesirable variations in 
performance. 

 

33B

Recommendation 

The duration of electrophoresis can be defined in time (hours/minutes). When using this or any 
electrophoresis product, we recommend that you adhere to the protocols given in the 
electrophoresis product manuals, and durations, specified in time. 

Important: 

For best results, follow these important guidelines when running multiple gels and 

electrophoresis units concurrently. For example: 

 

Avoid running samples of widely differing salt concentrations or sample buffers at the same 
time or on the same gel. 

 

Properly prepare your samples. 

Note: 

Variations in conductivity due to differences in buffer salt concentrations can affect the run 

of all the samples run at the same time. 

Note:

 For optimal performance, use Constant Voltage Mode when using the  

PowerEase

®

 300W Power Supply with 1 – 2 protein gels or blots. 

 

14B

Operation modes 

34B

Introduction 

The PowerEase

®

 300W Power Supply is designed to operate under different modes, 

Constant 

Mode 

or 

Program Mode

, designed to meet most electrophoresis needs. Use the 

Constant 

Voltage

,

 Constant Power 

or 

Constant Current Operation 

for applications that require only one 

specific voltage limit, current limit, or power limit continuously during the entire duration of 
electrophoresis. 

 

Содержание PowerEase PS0300

Страница 1: ...PowerEase 300W Power Supply Catalog Numbers PS0300 PS0301 PS0302 Publication Number MAN0008988 Revision D 0 Research Use Only Not for use in diagnostic procedures...

Страница 2: ...SE OF MERCHANTABILITY FITNESS FOR A PARTICULAR PURPOSE OR NON INFRINGEMENT TO THE EXTENT ALLOWED BY LAW IN NO EVENT SHALL LIFE TECHNOLOGIES AND OR ITS AFFILIATE S BE LIABLE WHETHER IN CONTRACT TORT WA...

Страница 3: ...peration modes 7 15BLimiting parameter settings 11 2BAppendix A 12 16BGeneral safety 12 Instrument safety 12 17BExplanation of symbols and warnings 13 1Product regulatory compliance 13 3BAppendix B 15...

Страница 4: ...ims must be filed with the carrier keep the supplied box for inspection The warranty does not cover in transit damage To ensure safe reliable operation always operate the PowerEase 300W Power Supply i...

Страница 5: ...wer Supply Constant Voltage Power or Current Capable of running multiple electrophoresis units LCD display with clear menu prompts for easy use Programming capabilities for limiting voltage V current...

Страница 6: ...e the cursor position down and decrease values when settings are flashing Mode Select Constant Mode or Program Mode The Mode key cycles between voltage current power and time settings allowing values...

Страница 7: ...riations in buffer conditions can result in exceeding the power supply s maximum voltage current or power output capacity and produce undesirable variations in performance 33BRecommendation The durati...

Страница 8: ...Mode the unit defaults to the Constant Voltage mode and Volts V displays on the left side of the screen in large characters The controlling value Volts V Amps A or Watts W is always displayed on the...

Страница 9: ...Power from the display screen The screen displays the selected operation mode on the left side of the display 5 Use the arrow keys to set the selected operation voltage VOLT current AMP or Power WAT t...

Страница 10: ...of the run press the Mode key 7 Voltage will appear on the display to enter voltage press Mode key 8 To change voltage use the arrow keys 9 Select AMP or WAT by pressing the Mode key until desired par...

Страница 11: ...ng margin of safety over time The same voltage setting can be used regardless of the number or thickness of gels being electrophoresed 42BCurrent limiting Discontinuous buffer systems and to a lesser...

Страница 12: ...indicator light at the Start Stop key is on or when RUNNING is displayed on the screen 2 WAIT at least 5 seconds after stopping a run before handling output leads or connected apparatus 3 ALWAYS make...

Страница 13: ...tri s lectif CAUTION Pour minimiser les consequences n gatives sur l environnement la suite de l limination de d chets lectroniques ne pas liminer ce d chet lectronique avec les d chets usuels non so...

Страница 14: ...uency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely...

Страница 15: ...cell Correct the condition by making sure the tape is removed from the pre cast gel buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit Operation s...

Страница 16: ...oved during a run Buffer leaking in a connected system Excessive temperature increase Excessive buffer evaporation Loose connection in a connected system Current set too low Change in constant mode Vo...

Страница 17: ...ghly conductive chloride ions in the gel are gradually replaced by the slower moving less conductive glycine ions from the running buffer as the gel runs As a result the resistance of the gel increase...

Страница 18: ...ant Current Constant Power 1V step 10 mA step 1 watt step Crossover Automatic Display type Backlit LCD Graphic type Display size 53 64 x 15 64 mm W x H Pause function Yes Safety features No Load Detec...

Страница 19: ...Appendix F To replace the fuse 1 Turn off the main power switch at the rear of the PowerEase 300W Power Supply and detach the power cord from the rear of the PowerEase 300W Power Supply 2 Open the fu...

Страница 20: ...tomer training Download software updates and patches 57BSafety Data Sheets SDS Safety Data Sheets SDSs are available at HUwww lifetechnologies com supportUH 58BCertificate of Analysis The Certificate...

Отзывы: