Orion Star A320 Series Portable Meter User Manual
Thermo Scientific
2
Meter Basics
Installing Batteries
Four AA batteries are included with Orion Star A320 series portable meters and are preinstalled
in the meter battery compartment. Use the following procedure to replace batteries as needed.
1.
Select four new AA alkaline batteries.
2.
Confirm that the meter is powered off.
3.
Turn the meter over, with the meter display facing down, on a clean dry surface.
4.
Loosen the two screws holding the
battery cover in place and gently
pull up the battery cover to release
it from the meter.
5.
Remove the old batteries from the
battery compartment.
6.
Install the new batteries in the
battery compartment, orientating
the batteries as shown in the
battery compartment housing.
7.
Replace the battery compartment
cover and tighten the two screws to secure the battery cover with the meter.
8.
Turn the meter over, with the meter display facing up, and power on the meter.
C H A P T E R 2
Loosen screws
Gently pull up to remove cover