Thermo Scientific Orion Star A212 Benchtop and Star A222 and Star A322 Portable Conductivity Meters
Reference Guide
Страница 1: ...Thermo Scienti c Orion Star A212 Benchtop and Star A222 and Star A322 Portable Conductivity Meters Reference Guide...
Страница 2: ...Triode Minimum Stir Requirement MSR NISS Optimum Results Orion Dual Star Orion Star SAOB SMART AVERAGING SMART STABILITY Star LogR Star Navigator 21 Stat Face Triode are trademarks of Thermo Fisher Sc...
Страница 3: ...ter Maintenance 8 Chapter 3 Meter Setup 9 Navigating the Setup Menu 9 Setup Menu Overview 9 COND Channel Menu 10 Settings Menu 12 View Log Menu 13 Diagnostics Menu 13 Chapter 4 Conductivity Calibratio...
Страница 4: ...ces 29 Meter Errors and Recommended Actions 29 Meter Self Test and Checkout 30 Meter Self Test 30 Meter Checkout Conductivity 31 Electrode Stability Test 32 Meter Factory Reset 32 Meter User Reset 32...
Страница 5: ...This page intentionally left blank...
Страница 6: ...OJUZ PS UP QQU B M FRVJWBMFOU XJUI QQU SFTPMVUJPO t 3FTJTUJWJUZ SBOHF PG PINT UP NFHB PINT XJUI PIN SFTPMVUJPO t 650 3 DPOUJOVPVT XJUI IPME PQUJPO BOE UJNFE NFBTVSFNFOU NPEFT t DBMJCSBUJPO FEJUJOH t Q...
Страница 7: ...GPVS TJHOJmDBOU mHVSF SFTPMVUJPO t 4BMJOJUZ SBOHF PG UP QQU QSBDUJDBM TBMJOJUZ PS UP QQU B M FRVJWBMFOU XJUI QQU SFTPMVUJPO t 3FTJTUJWJUZ SBOHF PG PINT UP NFHB PINT XJUI PIN SFTPMVUJPO t 650 3 DPOUJO...
Страница 8: ...DL PG UIF QPXFS BEBQUFS E POOFDU UIF QPXFS BEBQUFS UP UIF NFUFS BOE QPXFS PVUMFU BUUFSJFT JODMVEFE XJUI QPSUBCMF NFUFST TPME TFQBSBUFMZ GPS CFODIUPQ NFUFST a Select four AA alkaline batteries C POmSN...
Страница 9: ...UZ DFMM PS 5 QSPCF XJUI QJO JOJ DPOOFDUPS t SPVOE o OQVU GPS B HSPVOE XJSF XJUI B TUBOEBSE QJO UJQ DPOOFDUPS Orion Star A212 benchtop meters only t 1PXFS o OQVU GPS UIF QPXFS BEBQUFS t 4UJSSFS o OQVU...
Страница 10: ...E CBTF 2 Choose the side of the meter to attach the arm 6TF B DMFBO TVSGBDF BOE UVSO UIF NFUFS PWFS 3FMFBTF UIF FYJTUJOH TDSFX GSPN UIF CBDL PG UIF NFUFS MJHO UIF FMFDUSPEF BSN CBTF XJUI UIF DJSDMFT B...
Страница 11: ...EJDBUFT B NFUIPE JT JO VTF BOE UIF OVNCFS PG UIF NFUIPE CFJOH VTFE Cond OEJDBUFT UIF UZQF PG NFBTVSFNFOU BOE EFUFSNJOFT UIF UZQF PG DBMJCSBUJPO UIBU XJMM CF QFSGPSNFE ready Speci es the stability of t...
Страница 12: ...UIF DPOUJOVPVT NFBTVSFNFOU NPEF QSFTT UP IPME GSFF F UIF EJTQMBZFE NFBTVSFNFOU BOE QSFTT BHBJO UP SFMFBTF VOGSFF F UIF NFBTVSFNFOU O UIF TFUVQ DBMJCSBUJPO BOE PUIFS NFOVT QSFTT UP TDSPMM MFGU UISPVHI...
Страница 13: ...ZT BSF VTFE UP BDDFTT UIF TFUVQ NFOV DIBOHF UIF NFBTVSFNFOU NPEF NBOVBMMZ MPH PS QSJOU B NFBTVSFNFOU BOE IPME GSFF F B EJTQMBZFE NFBTVSFNFOU JO UIF continuous measurement mode O UIF DPOUJOVPVT NFBTVSF...
Страница 14: ...al place or negative sign press f3 enter UP TFMFDU UIF IJHIMJHIUFE JUFN BOE SFQFBU VOUJM UIF EFTJSFE WBMVF JT TIPXO PO UIF UPQ PG UIF OVNFSJD entry screen iii Press f2 done UP TBWF UIF WBMVF BOE FYJU...
Страница 15: ...conductivity cell used as standard or 641 4FMFDU TUBOEBSE GPS NPTU DPOEVDUJWJUZ DFMMT PS TFMFDU 641 GPS FMFDUSPEF MPX MFWFM DPOEVDUJWJUZ DFMMT FMM Default value or last entered value FOV VTFE UP FOUFS...
Страница 16: ...FSBUPS 1SFTT or to change the setting on each line press or to select another line and use the numeric entry screen to enter limit values Star A212 and Star A322 meters only Cal Due Alarm t BM VF MBSN...
Страница 17: ...GPSNBU GPS B QSJOUFS TUSJOH PG UFYU PS 1 DPNNB EFMJNJUFE UFYU BOE TFU UIF CBVE SBUF BUB PH t 0GG t 0O Menu used to turn on or off the data log feature BUF 5JNF t BUF t t t 5JNF t PVS MPDL t PVS MPDL F...
Страница 18: ...FS 6TF UIF BMQIBOVNFSJD FOUSZ TDSFFO UP FOUFS UIF VTFS Star A212 and Star A322 meters only 4BNQMF t 0GG t BOVBM t VUP ODSFNFOUBM FOV VTFE UP UVSO UIF TBNQMF PGG PS PO 8IFO TBNQMF JT TFU UP NBOVBM PS B...
Страница 19: ...14 Thermo Scienti c Orion Star A212 Benchtop and Star A222 and Star A322 Portable Conductivity Meter Reference Guide This page intentionally left blank...
Страница 20: ...tivity cell and standard are ready press f3 start G VTJOH B 4UBS CFODIUPQ NFUFS BOE TUJSSFS QSPCF UIF TUJSSFS QSPCF XJMM BVUPNBUJDBMMZ TUBSU TUJSSJOH BU UIF TQFFE TFMFDUFE JO UIF TFUVQ NFOV PODF f3 st...
Страница 21: ...press f2 print UP QSPDFFE UP UIF NFBTVSFNFOU NPEF BOE FYQPSU UIF EBUB UP B printer or computer or press f3 cal edit to edit another calibration point Manual Conductivity Calibration O UIF NFBTVSFNFOU...
Страница 22: ...ght Temperature Calibration and press f3 select 1SFTT or to highlight ATC UP JEFOUJGZ UIF 5 QSPCF UIBU XJMM CF DBMJCSBUFE BOE QSFTT f3 select 1MBDF UIF 5 QSPCF BOE UIFSNPNFUFST JOUP B TPMVUJPO XJUI B...
Страница 23: ...18 Thermo Scienti c Orion Star A212 Benchtop and Star A222 and Star A322 Portable Conductivity Meter Reference Guide This page intentionally left blank...
Страница 24: ...BOE XBJU GPS JU UP TUBCJMJ F B G UIF NFUFS JT JO 650 3 NPEF EFGBVMU TFUUJOH QSFTT to start the measurement When the JDPO TUPQT nBTIJOH SFDPSE UIF BQQMJDBCMF NFBTVSFNFOU QBSBNFUFST BOE UFNQFSBUVSF PG...
Страница 25: ...20 Thermo Scienti c Orion Star A212 Benchtop and Star A222 and Star A322 Portable Conductivity Meter Reference Guide This page intentionally left blank...
Страница 26: ...of each method 07 01 11 12 45 Cond if methods have been previously created 1SFTT or to highlight Current Settings and press f3 save 1SFTT or UP IJHIMJHIU B NFUIPE OVNCFS UIBU UIF DVSSFOU TFUUJOHT XJM...
Страница 27: ...ress or to highlight a number press f3 enter to select the highlighted OVNCFS BOE SFQFBU VOUJM UIF QBTTXPSE JT TIPXO c Press f2 done UP FYJU UIF OVNFSJD FOUSZ TDSFFO d Press f2 accept UP BDDFQU UIF FO...
Страница 28: ...E FOBCMFE JO UIF TFUVQ NFOV 3FWJFX UIF EBUF BOE UJNF TFUUJOHT BOE VQEBUF BT OFFEFE Read Type Selection O UIF NFBTVSFNFOU NPEF QSFTT 2 Press or to highlight COND Channel and press f3 select 3 Press or...
Страница 29: ...XJMM CF JOUFSGBDFE VTJOH UIF 64 QPSU 1SFTT or to highlight Data Format and press f3 select Press or to highlight Printer or PC CSV and press f3 select Select Printer UP FYQPSU UIF EBUB BT B TUSJOH PG...
Страница 30: ...tivity Resistivity TDS or Salinity and press f2 select 5IF NFUFS XJMM EJTQMBZ B MJTU PG DBMJCSBUJPOT GPS UIF TFMFDUFE DIBOOFM BOE DBMJCSBUJPO UZQF 5IF MJTU TIPXT UIF TFRVFOUJBM OVNCFS PG UIF DBMJCSBUJ...
Страница 31: ...FE and press f2 accept B G Today Last 10 or All is selected press f2 yes UP FYQPSU UIF EBUB C G Range JT TFMFDUFE o FOUFS UIF mSTU BOE MBTU EBUF QPJOUT i Press f3 edit to display the numeric entry scr...
Страница 32: ...t Conductivity Channel and press f3 select 5IF NFUFS XJMM EJTQMBZ B MJTU PG UIF EBUB QPJOUT 5IF MJTU TIPXT UIF TFRVFOUJBM OVNCFS PG UIF EBUB QPJOU BOE UIF EBUF BOE UJNF UIF EBUB QPJOU XBT TBWFE 07 01...
Страница 33: ...int a range of data log points or the entire data log Interfacing a Computer 5IF 0SJPO 4UBS BOE 4UBS TFSJFT NFUFST DBO TFOE NFBTVSFNFOUT BOE DBMJCSBUJPO EBUB UP B DPNQVUFS JO B DPNNB EFMJNJUFE GPSNBU...
Страница 34: ...s one at a time including the QPXFS LFZ XJUIJO GJWF TFDPOET PG POF PUIFS G UIF FSSPS QFSTJTUT DPOUBDU 5FDIOJDBM 4VQQPSU Over Range Measurement or 6OEFS 3BOHF FBTVSFNFOU G UIF NFBTVSFNFOU JT nBTIJOH BO...
Страница 35: ...FMFDUSPEFT BOE QSPCFT GSPN UIF NFUFS BOE QSFTT f2 yes 1SFTT f2 yes to start the self test 5IF NFUFS XJMM QFSGPSN BO BDDVSBDZ UFTU 8IFO UIF NFUFS EJTQMBZT Accuracy Test Passed press f2 next 5IF NFUFS X...
Страница 36: ...F BOE DPOEVDUBODF WBMVFT BSF QSJOUFE PO FBDI SFTJTUPS 5IF OPNJOBM SFTJTUBODF BOE DPOEVDUBODF WBMVFT BSF MJTUFE PO UIF MBCFM PG UIF DBMJCSBUJPO LJU CPY GPS SFGFSFODF 1 Remove the conductivity cell and...
Страница 37: ...JGU BOE OPJTF PO UIF EJTQMBZ 5IF NFUFS XJMM EJTQMBZ Stability Test Passed or Stability Test Failed Press f1 esc or to return to the measurement mode Meter Factory Reset GUFS B GBDUPSZ SFTFU JT QFSGPSN...
Страница 38: ...F EJTQMBZFE 1SFTT f1 back to return to the diagnostics menu or press to return to the measurement mode Assistance GUFS USPVCMFTIPPUJOH BMM DPNQPOFOUT PG ZPVS NFBTVSFNFOU TZTUFN DPOUBDU 5FDIOJDBM 4VQQP...
Страница 39: ...IUPQ FUFS XJUI MFDUSPEF 4UBOE STARA2145 4UBS Q 4 FODIUPQ FUFS XJUI MFDUSPEF 4UBOE 681 3044 6MUSB Q MFDUSPEF 5 1SPCF 4UJSSFS 1SPCF 3044 Q VGGFS JU STARA2146 4UBS Q 4 FODIUPQ FUFS XJUI MFDUSPEF 4UBOE 68...
Страница 40: ...FT 45 3 4 4UBS BSSZJOH BTF 45 3 3 4UBS FUFS SNPS STARA3230 4UBS 3 0 0 1PSUBCMF FUFS STARA3235 4UBS 3 0 0 1PSUBCMF FUFS 3 0 0QUJDBM 0 1SPCF XJUI FUFS BCMF 45 3 4 4UBS BSSZJOH BTF 45 3 3 4UBS FUFS SNPS...
Страница 41: ...BNUMD 3044 MBTT 3FGJMMBCMF Q 5 5SJPEF 8157BNUMD 3044 3FGJMMBCMF QPYZ Q 5 5SJPEF 8107UWMMD 3044 FM GJMMFE QPYZ Q 5 5SJPEF XJUI FUFS BCMF 927005MD QPYZ 4UFFM 5 1SPCF 927007MD Stainless Steel ATC Probe 9...
Страница 42: ...BQQBSBUVT EPFT OPU FYDFFE UIF MBTT MJNJUT GPS SBEJP OPJTF FNJTTJPOT GSPN EJHJUBM BQQBSBUVT TFU PVU JO UIF 3BEJP OUFSGFSFODF 3FHVMBUJPOT PG UIF BOBEJBO FQBSUNFOU PG PNNVOJDBUJPOT w i F QSFTFOU BQQBSFJM...
Страница 43: ...0SJPO 4UBS Q 0SJPO 4UBS Q POEVDUJWJUZ Orion Star A322 Conductivity 0SJPO 4UBS Q 3 0 0 0SJPO 4UBS 3 0 0 0SJPO 4UBS Q 4 0SJPO 4UBS Q POEVDUJWJUZ 0SJPO 4UBS Q 3 0 0 0SJPO 4UBS Q 4 POEVDUJWJUZ 3 0 0 Equi...
Страница 44: ...PSZ Weight 1PSUBCMF H FODIUPQ H Size 1PSUBCMF DN Y DN 8 Y DN FODIUPQ DN Y DN 8 Y DN Regulatory and Safety 567 MBTT 1PXFS 3BUJOH OQVU 7 BUUFSZ Y 4IPDL BOE 7JCSBUJPO 7JCSBUJPO TIJQQJOH IBOEMJOH QFS 45 4...
Страница 45: ...4UBS JOFBS UP O O 4UBS JOFBS UP O O O V 1 Compatible Cell Constants UP VNCFS PG BMJCSBUJPO 1PJOUT 6Q UP QPJOUT BMJCSBUJPO EJUJOH 4UBS FT 4UBS P 4UBS FT Resistivity Range PIN UP NFHB PIN Resolution PI...
Страница 46: ...EBUF TUBNQ 4UBS XJUI UJNF BOE EBUF TUBNQ PH VODUJPO 4UBS BOVBM 3FBEZ 650 3 5JNFE 4UBS BOVBM VUPNBUJD XJUI 650 3 4UBS BOVBM 3FBEZ 650 3 5JNFE PH EJU 4UBS FMFUF JOEJWJEVBM QPJOUT HSPVQ PG QPJOUT PS BMM...
Страница 47: ...42 Thermo Scienti c Orion Star A212 Benchtop and Star A222 and Star A322 Portable Conductivity Meter Reference Guide This page intentionally left blank...
Страница 48: ...scribed by the temperature coef cient of the solution The meter has a default temperature coef cient of 2 1 percent DIBOHF JO DPOEVDUJWJUZ QFS XIJDI JT SFQSFTFOUBUJWF PG NBOZ BRVFPVT TBNQMFT Solution...
Страница 49: ...tivity Standard Values vs Temperature Cat No 011005 01100510 011006 01100610 011007 01100710 01100910 011008 5FNQFSBUVSF N4 DN Conductivity Standard N4 DN N4 DN Conductivity Standard N4 DN 4 DN Conduc...
Страница 50: ...chtop and Star A222 and Star A322 Portable Conductivity Meter Reference Guide Cat No 011005 01100510 011006 01100610 011007 01100710 01100910 011008 31 112 32 33 121 123 131 212 222 Table of Conductiv...
Страница 51: ...This page intentionally left blank...
Страница 52: ...33 2463887 info water uk thermo com India Tel 91 22 4157 8800 wai asia thermo sher com Japan Tel 81 045 453 9175 wai asia thermo sher com China Tel 86 21 68654588 wai asi thermo sher com Singapore Tel...