Chapter 8
|
Customer Services
Meter Factory Reset Procedure
Warning: Performing a factory reset will erase the meter’s calibration log and data log and
reset all meter setup parameters to the factory default settings.
1.
In the measurement mode, press the setup key.
2.
Press the
p
,
q
,
t
or
u
key to highlight Diagnostics and press the f3 (select) key.
3.
Press the
p
or
q
key to highlight Factory Reset and press the f2 (select) key.
4.
Press the f2 (yes) key to start the factory reset procedure.
5.
Press the f3 (edit) key to access the popup number entry screen and enter the default
meter password of 111111.
a.
Press the
p
,
q
,
t
or
u
key to highlight the number 1 and press the f3 (enter) key
six times until 111111 is shown at the top of the screen.
b.
Press the f2 (done) key to save the value and exit the number entry screen.
6.
Press the f2 (accept) key to initiate the factory reset.
7.
Wait while the factory reset is performed. Once completed, the meter will go through a
power cycle and then proceed to the measurement mode. All meter settings will be reset to
the factory default settings and the calibration log and data log will be erased.
Meter User Reset Procedure
1.
In the measurement mode, press the setup key.
2.
Press the
p
,
q
,
t
or
u
key to highlight Diagnostics and press the f3 (select) key.
3.
Press the
p
or
q
key to highlight User Reset and press the f2 (select) key.
4.
Press the f2 (yes) key to initiate the user reset procedure.
5.
Wait while the user reset is performed. Once completed, the meter will proceed to the
measurement mode. All meter settings will be reset to the factory default settings; however
the calibration log and data log will be retained.
96
|
Orion Star A210 Series Benchtop Meter User Manual
Thermo Scientific