15
Scanning
Viewing and Manipulating Scan Data
72
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
3. Use
Cursor
→
and
Cursor
←
to position the left cursor line to the desired wavelength
value.
The instrument calculates the 3-point net absorbance for the selected wavelengths.
4. Continue selecting the other wavelengths by pressing
Next Cursor
to activate the center
and right cursor lines.
Select the wavelengths with
Cursor
←
and
Cursor
→
.
Repeat until all three wavelengths have been selected.
5. Press
Enter Factor
to access the set factor box. Enter the desired factor and press
Enter
.
The instrument calculates the value for the 3-point net absorbance for the selected
wavelengths, multiplied by the selected factor.
Calculating the Area Under a Curve
To calculate the area under a curve
1. With your scan data displayed on the edit graph screen, press
Math
.
2. Press
Area
to display the Area Under the Curve measurement screen.