14
Concentration Measurements—Advanced A-%T-C
Recalling a Test
62
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
Recalling a Test
To recall a test
1. Press
Stored Tests
.
2. Highlight the test and press
Enter
or
Load Test
to display the parameters for the selected
test.
Setting Up Test Parameters
To set up test parameters
1. Highlight the desired parameter.
Some parameters appear only if you select a concentration mode, while others appear
regardless of the selected measurement mode.
See
for a complete list.
2. When the parameters are set, press
Save Test
to save the test or
Run Standard
to measure
a standard (if in Conc/Std), or press
Run Test
(if in Conc/Factor).
Measuring a Standard
To measure a standard automatically (using Auto 6 or Auto 3)
1. With Measurement Mode set to Conc/Std, press
Run Standard
.
2. Enter the concentration of the standard and press
Enter
.
3. Press
Measure Blank
.