Water Analysis Application
Using a Reverse Color Method
46
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
Reverse color methods require that the instrument measure both a clear blank and a reagent
blank to properly calibrate the instrument before measuring test samples.
Using a Preprogrammed Reverse Color Method
Load the method from the library. An example of a reverse color method screen is shown
below.
To run the method, follow the sequence of steps below:
1. Press the
Run Test
function key. The following screen appears:
2. Place a cuvette, tube or vial containing a clear, colorless blank solution (such as distilled
water) in the measurement position and press the
Measure Blank
function key.
The
Measure Rgnt Blank
option will appear.