10
Water Analysis Application
Running a Test Method from the USB Stick
38
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
13. Press the
Run Test
function key to begin using the test method.
Refer to the specific reagent chemistry section for detailed instructions.
Running a Test Method from the USB Stick
1. Make sure the test method to be run is in the Thermo folder on the USB stick and the
USB stick is connected to the USB port on the front of the AquaMate
spectrophotometer.
2. From the initial measurement screen, press
test
to access the test types.
3. Press the
keys to highlight Water Analysis and press
enter
.
4. Press the
Stored Tests
function key.
5. Press the
keys to highlight Load test from USB drive and press
enter
.
6. Press the
keys to highlight the appropriate test method and press the
Load Test
function key.
7. Press the
Run Test
function key.
Running a Test Method from the AquaMate Instrument Library
1. From the initial measurement screen, press
test
to access the test types.
2. Press the
keys to highlight Water Analysis and press
enter
.
3. Press the
Stored Tests
function key.
4. Press the
keys to highlight Load internal test and press
enter
.
5. Press the
keys to highlight the appropriate test method and press the
Load Test
function key.
6. Press the
Run Test
function key.
Running a Test Method from the AquaMate SmartStart Menu
1. From the initial measurement screen, press
test
to access the test types.
2. Press the
SmartStart
function key.
3. Press the
keys to highlight the appropriate test method and press the
Load Test
function key.
4. Press the
Run Test
function key.