3
Accessories
External Printers
Thermo Scientific
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
17
Removing the 6-Position Cell Holder and the Single Cell Holder
To remove the 6-Position Cell Holder and the Single Cell Holder
1. Open the sample compartment door and let it rest on its hinge.
2. With one hand, loosen the captive thumbscrew.
3. With your other hand, pull straight up on the cell holder and lift it out of the sample
compartment
4. Close the sample compartment door.
Installing Accessory Cell Holders
Make sure that you have the correct cell holder baseplate installed.
See
for the different cell holder accessories that can
be created for your instrument.
Each of the Cell Holders needs to be installed on either a single-cell baseplate or a multi-cell
baseplate by removing the cell holder(s) secured to the baseplate. Each cell holder has a captive
screw in the bottom of the holder that secures the holder to the baseplate. Use a flat blade
screwdriver to loosen the captive screw from the baseplate and lift the cell holder from the
baseplate. Then insert the new cell holder into the appropriate position and secure it to the
baseplate by tightening the captive screw.
See
“Installing the 6-Position Cell Holder and the Single Cell Holder”
instructions on installing the complete accessory assembly in your instrument.
External Printers
Your spectrophotometer is able to print to external desktop printers supporting HP PCL 5.0
format and later.
Note
If the cell holder is not aligned correctly, you will not be able to tighten the
thumbscrews.
Note
To use 100 mm long pathlength cells, you must install the Single Cell Holder
baseplate.
Note
Remove the baseplate from the instrument before removing or installing cell
holders.
Note
PCL format does not support HP “Windows” printers.