3
Accessories
Cell Holders and Cell Holder Accessories
14
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
Cell Holder Initialization
When you press
Run Test
to start a measurement, the instrument displays the message
“Calibrating and Checking Turret, Please Wait” while it attempts to initialize the cell changer
at the Blank position. With the Water Analysis carousel, the spectrophotometer will rotate the
carousel to the default position, then rotate the selected cell holder back into the measurement
position. This ensures that the cell holder is properly positioned in the beam.
If a single cell holder has been installed, the message “Error, Single Cell Holder found. Use
Single Cell Holder?” appears. Press
Accept Change
to continue, or install the cell changer and
press
Cancel Change
.
Cell Changer System
Single Cell System
Standard Cell Holders