Menu Overview
Section 7
2230XP Silica Analyzer
Thermo Scientific Orion
53
NOTE:
To enter the Service menu, the instrument must be in manual mode. (Press
STOP.)
Service Menu
System Administration Menu
Allows access control to the instrument for up to six users and one technician, all with individual
passwords.
Technician alone has the permission to add, delete and set access permission of any user.
Each user can have access to edit their own user name and password.
Menus also allow technician to change factory settings (factory password required), printing
options, instrument label, and to view instrument setup details.
Hardware Setup Menu
Allows technician to calibrate several hardware components of the analyzer.
Calibration results are editable by the technician.
Prime Fluidics Menu
Allows user to prime the fluidics lines of the analyzer.
Diagnostics Menu
Diagnostics provide a direct mechanism to individually test devices such as pumps, valves,
relays, optics and stirrer. Turn various pumps and valves ON and OFF for testing or during
adjustment. Maximum of two devices can be turned on at the same time. This function also
provides a quick check of the optics subsystem.
System Administration Menu
Edit User Menu
Allows technician to edit user names, user passwords and user permissions, for measurement,
calibration and validation. Allows user to edit his or her own user name and password.
Delete User Menu
Allows technician to delete a user from the users list.
Add User Menu
Allows technician to add a new user to the user’s list. Up to six different users may be added.