Electrical Connections
14
Orbitrap Fusion Series Getting Connected Guide
Thermo Scientific
To connect the MS and data system cables and power cords
1
1
The back of the data system computer might not be as shown.
(Not shown) Connect both ends of the
power cords
(for the computer, Ethernet switch, modular valve,
syringe pump, MS, and forepumps).
IMPORTANT
Plug the power cords for the MS and
forepumps into separate, dedicated 230 Vac electrical
outlets—that is, circuits A, B, and C for a
dual-forepump system.
Note
If the modular valve’s LED remains off, press its
front Remote button to exit local mode.
5
1
Connect the
relay control
cable
between the MS
(V65 Pump Enable) and the
forepumps.
4
(Not shown) Connect the
contact
closure cable
and the
Ready Out
cable
(not provided) between the
MS and the external device. See
Chapter 6, “External Devices.”
2
Connect the
USB cables
from the
MS to the syringe pump and the
modular valve.
3
Connect the
Ethernet cables
from
the Ethernet switch to the computer’s
LC/MS Ethernet network card and the
MS.
IMPORTANT
Before turning on the MS, make sure to complete all of the system
connections described in this guide.
Tip
For troubleshooting purposes, you might want to record which devices connect to
which ports on the Ethernet switch.
Содержание Orbitrap Fusion Lumos MS
Страница 1: ...Orbitrap Fusion Series Getting Connected Guide 80011 97002 Revision C June 2017...
Страница 6: ......
Страница 8: ...Contents viii Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...
Страница 24: ...4 Syringe Pump and Modular Valve Assembly 12 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...
Страница 28: ...5 Electrical Connections 16 Orbitrap Fusion Series Getting Connected Guide Thermo Scientific...