Installation
Installing the Stacking Feet
3-10
Heratherm
Thermo Scientific
Stacking Kit
The stacking adapter is available only for table-top units.
Scope of Delivery:
1 stacking adapter
1 anti-tilt anchor
1 plastic bag with 2 stacking feet and 2 M4x16 Torx screws.
Required Tools:
Slotted screwdriver 5,5x100 or Torx screwdriver 20x100.
Installing the Stacking Feet
Remove the left and right blank plugs at the top blank.
Install the stacking feet with the enclosed screws using a slotted or Torx screwdriver.
Material Number
Description
50126665
Stacking adapter Heratherm 60L
50126666
Stacking adapter Heratherm 100L
50126667
Stacking adapter Heratherm 180L
Содержание OGS 100
Страница 1: ......
Страница 6: ...Contents iv Heratherm Thermo Scientific...
Страница 46: ...Product Description Tube Access Ports 4 16 Heratherm Thermo Scientific...
Страница 60: ...Start up Connecting the RS 232 Interface 5 14 Heratherm Thermo Scientific...
Страница 76: ...Handling and Control Switching the Oven Off Powering Down 7 14 Heratherm Thermo Scientific...
Страница 78: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific...
Страница 82: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific...
Страница 88: ...Maintenance Returns for Repair 10 6 Heratherm Thermo Scientific...
Страница 104: ...Spare Parts and Accessories 14 6 Heratherm Thermo Scientific...
Страница 106: ...Device Log 15 2 Heratherm Thermo Scientific...
Страница 108: ...Contact 16 2 Heratherm Thermo Scientific...
Страница 109: ...Contact Thermo Scientific Heratherm 16 3...
Страница 110: ...Contact 16 4 Heratherm Thermo Scientific...