![Thermo Scientific OGS 100 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/thermo-scientific/ogs-100/ogs-100_operating-instructions-manual_1105331019.webp)
Installation
Room Ventilation
Thermo Scientific
Heratherm
3-3
Room Ventilation
Heat dissipating from the oven during continuous operation may cause a change in the room
climate.
• Therefore, the oven must only be installed in rooms with sufficient ventilation.
• Do not install the oven in room recesses without ventilation.
• When several devices are to be placed in the same room, additional ventilation may have to
be provided as necessary.
• To avoid any impact of the heat dissipated by the oven on the ambient climate the room
must be vented by means of a laboratory-grade ventilation system that complies with
applicable local and national health and safety regulations and has sufficient capacity.
• If excessive temperatures tend to occur in the operating room, be sure to provide a thermal
protection means that cuts out the power supply to mitigate the impact of
overtemperature scenarios.
Use with Air Exhaust Systems
When the oven is connected to an air exhaust system be sure to adjust its air flow so the
temperature distribution pattern throughout the oven’s workspace remains unchanged and
precision temperature control is maintained.
Before using existing piping in the building to exhaust hot air from the oven, be sure that such
exhaust piping is made of heat-resistant material, so it will not overheat, melt and/or cause a
fire.
The oven’s outer surfaces and its air exhaust piping may become hot, so be sure to maintain the
proper spacing from walls and ceilings specified by local building codes and/or fire protection
regulations.
• Mark hot exhaust air piping with appropriate hot surface warning signs or restrict access to
such piping altogether to avoid the risk of injury through contact with hot surfaces.
Содержание OGS 100
Страница 1: ......
Страница 6: ...Contents iv Heratherm Thermo Scientific...
Страница 46: ...Product Description Tube Access Ports 4 16 Heratherm Thermo Scientific...
Страница 60: ...Start up Connecting the RS 232 Interface 5 14 Heratherm Thermo Scientific...
Страница 76: ...Handling and Control Switching the Oven Off Powering Down 7 14 Heratherm Thermo Scientific...
Страница 78: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific...
Страница 82: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific...
Страница 88: ...Maintenance Returns for Repair 10 6 Heratherm Thermo Scientific...
Страница 104: ...Spare Parts and Accessories 14 6 Heratherm Thermo Scientific...
Страница 106: ...Device Log 15 2 Heratherm Thermo Scientific...
Страница 108: ...Contact 16 2 Heratherm Thermo Scientific...
Страница 109: ...Contact Thermo Scientific Heratherm 16 3...
Страница 110: ...Contact 16 4 Heratherm Thermo Scientific...