Radiation and General Safety
EMERGENCY PROCEDURES
3-16
NITON XL3 Analyzer User’s Guide
Thermo Scientific
EMERGENCY
PROCEDURES
THIS PAGE CONTAINS EMERGENCY CONTACT INFORMATION
THAT SHOULD BE AVAILABLE TO THE OPERATOR AT ALL
TIMES.
Lost or Stolen Instrument
If the NITON XL3t analyzer is lost or stolen, notify your Radiation Safety
Officer (RSO) or the equivalent responsible individual at your company or
institution immediately. Your company's RSO, as well as other important
emergency contacts, are listed below. Your company RSO may need to
notify the x-ray tube regulatory authority and the local police. It is also
recommended that a notification is made to Thermo Fisher Scientific.
Damaged Instrument
Minor Damage
If the instrument is intact but there is indication of an unsafe condition such
as a cracked case, a shutter mechanism failure, or the lights remain flashing
after a measurement is terminated, follow these steps:
1. Stop using the instrument
2. Remove the battery, the x-ray tube can not produce radiation
when the battery is disconnected. The instrument is now safe to
handle.
3. Place the instrument securely in the holster
4. Place the instrument in the carrying case that came with the
instrument.
5. Notify your Radiation Safety Officer (RSO) or the equivalent
responsible individual at your company or institution
immediately.
6. You or your RSO should call Thermo Fisher Scientific at one of
their contact numbers listed below for additional instructions
and guidance.
Содержание NITON XL3t 500
Страница 1: ...NITON XL3t 500 Analyzer User s Guide Version 6 5 ...
Страница 2: ......
Страница 22: ...Applications Live Spectrum Feed 1 8 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Страница 40: ...The Measurement Screen 1 26 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Страница 66: ...The View Data Screen 1 52 NITON XL3 Analyzer User s Guide Thermo Scientific Figure 1 8 The SpectraView Screen ...
Страница 120: ...The Sort Element Display Menu 1 106 NITON XL3 Analyzer User s Guide Thermo Scientific Figure 1 40 Save Changes ...
Страница 160: ...Entering Data with a Barcode reader 2 32 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Страница 182: ...Radiation and General Safety Contact Information 3 22 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Страница 200: ...Appendix F Appendices xviii NITON XL3 Analyzer User s Guide Thermo Scientific ...