Maintenance
Calibrate the Laser Frequency
Calibrate your laser frequency if a qualification workflow, such as the Nicolet FTIR - Factory
Qualification workflow, indicates a failing test. When a test fails, first align the spectrometer, then
calibrate the laser frequency and run the test again. If the test still fails, contact your local technical
support representative for help.
❖
To calibrate the laser (touchscreen interface)
1. On the home screen, select the Diagnostics icon [
] to open the Diagnostics view.
2. Open the Laser Calibration tab and select Calibrate to begin.
When laser calibration is complete, a message indicates whether the laser frequency has been
changed and displays the new frequency.
❖
To calibrate the laser (desktop interface)
1. Select Acquire Data > Diagnostics > Laser Calibration.
2. In the Laser Calibration dialog box, click Start to calibrate the laser.
When the system is finished calibrating the laser, a message indicates whether the frequency was
changed and displays the new frequency.