Replace the Source

The IR source is easily accessible from the bottom of the spectrometer and can be replaced 
without having to remove the spectrometer’s cover.

Tools and materials needed

• A #1 Phillips head screwdriver

• Source Replacement Kit

Time needed: 25 minutes or less

To replace the source

1. Turn off and disconnect the spectrometer.

a. Turn off the instrument and unplug the power cord.

b. Disconnect any cables from the instrument, such as the Ethernet cable or USB 

devices.

c.

If the spectrometer or an installed accessory is purged, disconnect the purge lines 
from the spectrometer and the accessory (quick connect fittings stop the flow 
automatically). See 

Install and Maintain a Purge Kit 

for more information.

d. Remove any installed accessories from the instrument. 

CAUTION  

Before replacing the source, turn off the instrument. Replace the source only 

with replacement parts supplied by us.

CAUTION  

The source becomes extremely hot while you are using the instrument. 

Wait at least 10 minutes after turning the instrument off before you perform the 
next step.

Содержание Nicolet Summit PRO

Страница 1: ...ny cables from the instrument such as the Ethernet cable or USB devices c If the spectrometer or an installed accessory is purged disconnect the purge lines from the spectrometer and the accessory quick connect fittings stop the flow automatically See Install and Maintain a Purge Kit for more information d Remove any installed accessories from the instrument CAUTION Before replacing the source tur...

Страница 2: ...screen monitor attached move the monitor to the storage position as show in Figure 1 Figure 1 Summit PRO spectrometer with touchscreen monitor in the storage position b Carefully tip the instrument back until it rests on the fins of the heat sink Do not use the touchscreen monitor to support the weight of the spectrometer ...

Страница 3: ...he instrument b Tighten the screws to secure the source and snap the bale into its storage position 4 Reconnect and power on the spectrometer a Carefully return the instrument to its upright position and reconnect any cables you removed b Replace the desiccant See Replace the Desiccant for instructions Because the internal optics are exposed to the atmosphere when removing the source the desiccant...

Страница 4: ...navigate to the Workflows tab ii Select the Nicolet FTIR Factory Qualification workflow and touch the Run icon Follow the on screen prompts to complete the workflow To use the desktop interface perform the following steps a Align the spectrometer i Open OMNIC Paradigm software ii Open the Acquire Data menu and select Diagnostics Align Spectrometer iii Click Start and follow the on screen prompts C...

Страница 5: ...Replace the Source Thermo Scientific Nicolet Summit User Guide 5 2019 Thermo Fisher Scientific Inc All rights reserved 269 334400_Revison A ...

Отзывы: