iS50 ATR Module
8
iS50 ATR Module User Guide
Thermo Scientific
The extra tip can be stored on the pressure arm.
Removing and Installing the Sampling Plate
Figure 5.
Sampling plate features
Move the pressure arm to the sampling position (shown above) before you remove the
sampling plate.
To release the sampling plate, lift the back edge. To reinstall it, gently press down on the plate
until it seals against the O-ring around the crystal plate.
Table 4.
Tip styles
Tip style
Use with...
Flat
The end of this tip is flat, which makes it ideal for measuring films
and flat plastics as well as tiny shavings and fibers. This is a general
purpose tip style.
Concave
The end of this tip is slightly concave to allow good contact with
powders and curved solids such as a tablet or polymer bead.
NOTICE
• Make sure the sampling plate is installed before you place a sample on the ATR
crystal.
• Avoid damaging or contaminating the ATR internal optics. Keep samples and other
lab materials away from the ATR module when the sampling plate has been removed.
Sampling plate