NanoDrop One User Guide
29
Setting
Available Options
Description
Sample type and
Factor
dsDNA (with non-editable factor of
50 ng-cm/μL)
Widely accepted value for double-stranded DNA
ssDNA (with non-editable factor of
33 ng-cm/μL)
Widely accepted value for single-stranded DNA
RNA (with non-editable factor of
40 ng-cm/μL)
Widely accepted value for RNA
Oligo DNA with non-editable calculated
factor in ng-cm/μL
Factor calculated from user-defined DNA base sequence.
When selected, available DNA base units (i.e., G, A, T,
C) appear as keys. Define sequence by tapping
appropriate keys. Factor is calculated automatically based
on widely accepted value for each base unit.
Oligo RNA with non-editable calculated
factor in ng-cm/μL
Factor calculated from user-defined RNA base sequence.
When selected, available RNA base units (i.e., G, A, U,
C) appear as keys. Define sequence by tapping
appropriate keys. Factor is calculated automatically based
on widely accepted value for each base unit.
Custom (with user-specified factor in
ng-cm/μL)
Enter
between 15 ng-cm/μL and 150 ng-cm/μL
Dye 1/Dye 2 Type
a
a
To add a custom dye or edit the list of available dyes, use the Dye/Chromophore Editor.
Cy3, 5, 3.5, or 5.5,
Alexa Fluor 488, 546, 555, 594, 647, or
660
Select pre-defined dye(s) used to label sample material,
or one that has been added using Dye/Chrom. Editor.
Dye 1/Dye 2 Unit
picomoles/microliter (pmol/uL),
micromoles (uM), or millimoles (mM)
Select unit for reporting dye concentrations
Analysis
Correction
b
b
The Analysis Correction affects the calculation for nucleic acid concentration only.
On or off
Enter analysis correction wavelength in nm
or use default value (340 nm)
Corrects sample absorbance measurement for any offset
caused by light scattering particulates by subtracting
absorbance value at specified analysis correction
wavelength from absorbance value at analysis
wavelength. Corrected value is used to calculate sample
concentration.
Tip
: If the sample has a modification that absorbs light
at 340 nm, select a different correction wavelength or
turn off Analysis Correction.
Содержание NanoDrop One
Страница 1: ...NanoDrop Micro UV Vis Spectrophotometers NanoDrop One User Guide 269 309102 Revision A May 2017 ...
Страница 18: ......
Страница 28: ......
Страница 58: ......
Страница 134: ...13 Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Страница 144: ......
Страница 152: ......
Страница 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Страница 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Страница 193: ...18 Learning Center Set Up the Instrument Thermo Scientific NanoDrop One User Guide 187 Instrument Settings ...
Страница 293: ...18 Learning Center Multimedia Thermo Scientific NanoDrop One User Guide 287 Pedestal Cleaning and Reconditioning ...
Страница 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Страница 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Страница 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...