Maintenance
Regular and preventive maintenance
74
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
Figure 5–27.
Removing the reagent filter with tube
l
Place a disposable reagent basin filled with
detergent
, for example,
filtered 1% Micro-90 or equivalent, under the dispensing valves so
that the liquid is in contact with the valves (Figure 5–24).
l
Press the
EMPTY
button for several minutes.
l
Place another disposable reagent basin now filled with
distilled and
filtered laboratory-grade water
under the dispensing valves so that the
liquid is in contact with the valves (cf. Figure 5–24).
l
Press the
EMPTY
button for several minutes.
3.
Clean the reagent reservoir.
4.
Replace the 43 µm reagent filter and the tube into the cap of the
reagent reservoir.
5.
Add distilled and filtered laboratory-grade water into the reagent
reservoir. Close the cap.
6.
Press the
PRIME
button to rinse the tubing.
7.
Try to dispense.
8.
Repeat this entire procedure, Steps 2 to 7, if the dispensing valve is
still leaking.
Now the instrument is ready for dispensing.
Содержание Multidrop Combi nL
Страница 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Страница 98: ...Certificate of Decontamination 98 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Страница 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...