Installation
Requirements
18
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
When you set up your Multidrop Combi nL, avoid sites of operation
with excess dust, vibrations, strong magnetic fields, direct sunlight or
UV light, draft, excessive moisture or large temperature fluctuations.
l
Make sure the working area is flat, dry, clean and vibration-proof
and leave additional room for accessories, cables, and reagent
bottles.
l
Make sure the ambient air is clean and free of corrosive vapors,
smoke and dust.
l
Make sure the ambient temperature range is b10°C (50°F)
and +40°C (104°F).
l
Make sure the humidity is so low that condensing does not occur
(relative humidity is between 10% and 80%).
Leave sufficient space (at least 10 cm) on both sides and at the back of
the unit to allow adequate air circulation.
The Multidrop Combi nL does not produce operating noise at a level
that would be harmful. No sound level measurements are required after
installation.
Place the instrument on a normal laboratory bench. The net weight of
the entire equipment is approx. 9.6 kg (21.2 lbs.).
The instrument operates at voltages of 100–240 Vac and a frequency
range of 50/60 Hz.
l
Always ensure that the local supply voltage in the laboratory
conforms to that specified on the rating label on the back of the
instrument (Figure 2–4).
l
Do not smoke, eat or drink while using the Multidrop Combi nL.
l
Wash your hands thoroughly after handling test fluids.
l
Observe normal laboratory procedures for handling potentially
dangerous samples.
l
Wear proper protection clothing, such as disposable gloves and
laboratory coats, according to good laboratory practice.
l
Ensure that the working area is well ventilated.
l
Never spill fluids in or on the equipment.
This section describes the installation setups that you must carry out
before operating or relocating the instrument.
Requirements
Precautions and
limitations
Installation
setups
Содержание Multidrop Combi nL
Страница 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Страница 98: ...Certificate of Decontamination 98 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Страница 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Страница 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...