Operation
Calibrating the Micro-Tech
Thermo Fisher Scientific
Micro-Tech 9101/9201 Reference Manual, Rev J
3-7
3.
Press the Zero Calibration button and the following screen appears.
4.
Make sure the belt is running empty and at maximum speed, then
press the Start button. The count-down screen appears. (The data in
your screen will, of course, be different.)
NOTE.
The number of seconds shown in “Time remaining” is
calculated based on the current speed-sensor pulse frequency, and
estimates the time remaining for a complete test.
5.
The calibration time (in seconds) that you established during the
Micro-Tech cold-start procedure, will start counting down. When the
counter reaches zero, the calibration is complete and the change-zero
screen appears.
BATCH
ZERO CAL
Run the belt empty, then
press START
READY
ALARM
CALIB
EXIT
START
MANUAL
BATCH
AUTO ZEROING
Time remaining 375
Rate 0.00 Tph
Tot 0.000 Tons
READY
ALARM
CALIB
ABORT
Содержание Micro-Tech 9101
Страница 1: ...Micro Tech 9101 9201 Integrator Reference Manual REC 4281 Rev J Part Number 127336 English...
Страница 82: ......
Страница 118: ......
Страница 124: ...A 2 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific...
Страница 172: ......
Страница 234: ......
Страница 315: ......
Страница 316: ......
Страница 317: ......
Страница 318: ......
Страница 319: ......
Страница 320: ......
Страница 321: ......
Страница 322: ......
Страница 323: ......
Страница 324: ......
Страница 325: ......
Страница 326: ......
Страница 327: ......
Страница 328: ......
Страница 329: ......