Tuning and Calibrating Automatically in the ESI/MS Mode
Setting Up the Syringe Pump for Tuning and Calibration
46
LTQ XL Getting Started
Thermo Electron Corporation
a. Connect the infusion line with a (brown) fingertight fitting and a
(brown) ferrule to the free end of the LC union.
b. Connect the other end of the infusion line with a (red) fingertight
fitting and a (brown) ferrule to the grounding union.
The syringe pump is now properly set up for infusing solution into the MS
detector.
Go to the next section,
Setting Up the MS Detector in the Xcalibur Data
System for Tuning and Calibration
Figure 17.
ESI/MS plumbing connections for the fused-silica infusion line